Primary Service
CMM Calibration and Repair
Request a quotation for calibration or repair.

Request a quotation
Verification, not a label
Calibration establishes a measured relationship with a stated uncertainty. What most machines need, and what gets booked under that name, is performance verification against ISO 10360-2 or ASME B89.4.1 — confirming the machine still meets its stated specification.
Each frame type carries its own test standard: a bridge or gantry machine is verified under ISO 10360-2, a portable arm under ISO 10360-12 or ASME B89.4.22, and a rotary table fourth axis under ISO 10360-3, where applicable to the configuration on site.

The sequence of a visit
Each visit runs in the same order regardless of frame type, starting with the parts of the machine that drift fastest.
- 01
Scales and reader heads
X, Y and Z scales are inspected and cleaned; reader head output is checked before geometry testing.
- 02
Air bearings and airlines
Bearing condition and supply pressure are checked; contamination here shows up as noise in the length test.
- 03
Geometry
Straightness, linearity and squareness are measured in the XY, ZX and ZY planes.
- 04
Length and probing error
ISO 10360-2 length measurement error at 105 measurements, and probing form error, designated PFTU under ISO 10360-5, measured at 25 points on a dedicated test sphere.
As-found to as-left
- 01
As-found test
Results are captured against the stated MPE before any adjustment is made.
- 02
Adjustment
Geometric errors are corrected and, where the controller allows it, the volumetric compensation map is updated.
- 03
As-left verification
The length and probing tests are repeated to confirm the machine meets specification.
- 04
Certificate issue
Results, uncertainty, method and traceability are issued under the performing laboratory's scope.

Why Memphis CMM Calibration and Repair
- 01
As-found recorded first
Adjustment never happens before the as-found data is on record.
- 02
Repair alongside calibration
Probe heads, scales, air bearings and drives are serviced without a second vendor visit.
- 03
Frame-appropriate standard
Arms, trackers and rotary tables are tested where applicable to that configuration, not treated as bridge-machine extensions.
Coverage across West Tennessee and the Mid-South
On-site scheduling and ship-in service are available across the Mid-South; travel terms and radius are confirmed at quote.
Other services
Discuss CMM Calibration and Repair Requirements
Scope a calibration or repair visit now.